Biocompatibility assessment is required for any medical device that contacts the patient — directly or indirectly. FDA's expectation is that biocompatibility be evaluated using the ISO 10993 series of standards. Biocompatibility gaps are among the most common reasons 510(k)s receive AI requests, and they are almost entirely avoidable with thorough preparation.

Does your device need biocompatibility assessment?

Any device that contacts the patient — including indirect contact through circulating blood or bodily fluids — requires biocompatibility evaluation. This includes devices that contact intact skin, mucosal membranes, breached skin, blood, tissue, bone, or dentin. External communicating devices that contact blood or body fluids externally also require evaluation.

Non-contacting devices (e.g. imaging equipment that doesn't touch the patient, software-only devices) do not require biocompatibility assessment.

ISO 10993-1 CONTACT NATURE AND DURATION MATRIX
Contact Type
Limited (<24h)
Prolonged (24h–30d)
Permanent (>30d)
Surface contact (skin)
Cytotoxicity, sensitization
+ Irritation
+ Systemic toxicity
Mucosal membrane
Cytotoxicity, sensitization, irritation
+ Subacute tox
+ Subchronic tox, genotoxicity
Breached/compromised
Cytotoxicity, sensitization, irritation, acute tox
+ Subacute, genotoxicity
+ Implantation, chronic tox
Blood contact
Cytotoxicity, sensitization, hemocompatibility, pyrogenicity
+ Subacute, genotoxicity
+ Chronic, carcinogenicity
Tissue/bone/dentin
Cytotoxicity, sensitization, irritation, acute tox
+ Subacute, genotoxicity, implantation
+ Chronic, carcinogenicity

The ISO 10993-1 risk-based approach

ISO 10993-1 (updated in 2018) requires a risk-based approach to biocompatibility. Rather than automatically running every test in the matrix, you must first perform a biological evaluation — a documented assessment of your device's materials, manufacturing processes, and contact characteristics — and use that assessment to determine which tests are needed, which can be justified by existing data, and which can be waived.

FDA's guidance on the use of ISO 10993-1 (issued 2016, updated 2023) aligns with this approach and specifically states that manufacturers should not default to running all tests in the matrix. Over-testing without a documented rationale is as problematic as under-testing.

When literature justification is acceptable

If your device uses materials with an established biocompatibility history — such as materials that have been used in many cleared predicate devices — you may be able to justify some endpoints with literature or existing data rather than new testing. This is particularly common for:

  • Standard medical-grade polymers (silicone, polyurethane, PEEK) with extensive published biocompatibility data
  • Metals with established biocompatibility records (titanium, stainless steel, cobalt-chrome alloys)
  • Materials that have been used in cleared predicate devices with the same contact type and duration

Literature justification must be documented and specific — citing a general reference that the material is "commonly used" is not sufficient. The reference must cover the relevant endpoints for your contact type and duration.

Common biocompatibility AI triggers

  • Missing endpoints — failing to address one or more endpoints required for your contact type and duration
  • Wrong contact duration — testing for limited contact when your device is actually prolonged or permanent use
  • Processing residuals not addressed — manufacturing aids (mold release agents, lubricants, cleaning agents) that may leave residuals must be addressed even if the base material is biocompatible
  • Colorants and additives not evaluated — colorants, stabilizers, and other additives in plastics require separate biocompatibility consideration
  • New materials not tested — citing a predicate device's biocompatibility when your device uses different or additional materials
  • Outdated test standards — testing to superseded versions of ISO 10993 part standards
Find predicates with matching materials

Search FDA 510(k) clearances to find predicates using the same materials and contact configuration — so you can leverage existing biocompatibility data.

Start free trial →
Related articles
FDA 510(k) Submission Checklist: Every Section You Need → FDA 510(k) Labeling Requirements → How to Find a Predicate Device for Your 510(k) →

Everything you need for a 510(k) submission. Nothing you don't.

14-day free trial. No setup. Cancel anytime.

Start free trial →
Building an AI or ML-enabled device? Predicate search, PCCP tracking, and AI-specific FDA intelligence — built exclusively for AI/ML devices. Try AIFDA Intel →