K-numberK244054
Device name16ch Flex SPEEDER Medium 1.5T (Q7000246); 16ch Flex SPEEDER Large 1.5T (Q7000247)
ApplicantQuality Electrodynamics
Product codeMOS
Device classClass II
Decision dateFeb 28, 2025
DecisionSubstantially Equivalent
Regulation892.1000
AI Summary extracted from FDA summary PDF · never regenerated
Intended use

The 16ch Flex SPEEDER Medium and Large are receive-only, 16-channel phased array coils designed for magnetic resonance imaging on Canon 1.5T MR systems. They are intended to produce diagnostic images of general human anatomy that can be interpreted by a trained physician.

Technological characteristics

Both the proposed and predicate devices are based on identical technological elements: receive-only phased array RF coils with active PIN diode switching and passive blocking circuitry, housed in polycarbonate with polyurethane-coated nylon fabric covers for impact protection and biocompatibility. No technological differences exist between the proposed and predicate devices.

Test standards cited

Signal-to-noise ratio (SNR) and image uniformity were analyzed per NEMA MS-9 (using alternate method 2.5 from MS-6) and NEMA MS-9 (primary method from MS-6). Electrical safety and electromagnetic compatibility testing followed IEC standards. Surface heating was tested per AAMI/ANSI EN60601-1.

Substantial equivalence argument

The proposed devices have identical technological design, materials, and construction to the predicate device. The indications for use differ only in scope—the proposed devices cover general human anatomy while the predicate specified specific anatomies—but both are used with Canon 1.5T systems to produce diagnostic images. Testing demonstrates equivalent or better performance: SNR and uniformity meet acceptance criteria, surface heating remains below 41°C, clinical images from multiple anatomical regions show diagnostic quality, and no adverse events were observed.

Extracted by AI from the official FDA summary PDF →
Source

View the full FDA submission: accessdata.fda.gov

Researching this as a predicate?
Want a transparent AI-ranking score, AI-discovered related predicates, ongoing safety and warning-letter monitoring, full predicate chain lineage, and a drafted SE rationale — all saved to your own project? That's what an account adds.
Start free trial →

Everything you need for a 510(k) submission. Nothing you don't.

14-day free trial. No setup. Cancel anytime.

Start free trial →
Building an AI or ML-enabled device? Predicate search, PCCP tracking, and AI-specific FDA intelligence — built exclusively for AI/ML devices. Try AIFDA Intel →